3
Important Start-up Instructions
(Follow these instructions prior to using the system)
1. Attach the AC Power Cord to the power inlet on the Sapphire Plus light source and plug into AC outlet.
2. Install the Curing, Lesion Detection or Soft Tissue Management (STM) device into front panel receptacle
(black adapter insert must be removed to install STM). Make sure the device is firmly inserted all the way into
the receptacle before turning on the power. If the device is not installed correctly the system display will show
“no HP” meaning No Hand Piece is detected.
3. When using the built-in light meter, the handpiece should be held in the position as indicated in the illustration.
The tip should be centered inside of the aperture. Positioning the tip differently may result in reduced light
intensity values.
4. Do not saturate the controller keypad area with disinfectant or other liquids. Wipe the keypad surface with a
damp towel and dry completely.
5. If the lamp access plate on the bottom of the system is removed to check
the lamp, make sure that it is replaced correctly. The corners of the plate
must fit into the outline of the cutout area and not rest on top of the edges of
the surrounding metal. When installed properly all surface areas are even with
each other. If the access plate is misaligned, the lamp will not activate. If the
access plate is misaligned the system will show “Locd” meaning system is
Locked and the lamp will not activate.
Dangerous Voltage
Attention: consult accompanying documents
Attention: may be hot
Protection Class BF, Insulation Class 1
Alternating Current
Fuse
Enclosed Equipment Protected
Against Dripping Water
Hazard, High-Intensity Light
(applies when Curing/Whitening
Handpiece is attached)
DO NOT use in the presence of
flammable substances
Laser Radiation – Avoid eye
or skin exposure to direct or
scattered radiation (applies
when Laser Handpiece
is attached)